ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2012-13520
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Report Date
- October 31, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(4).
(B)(4): ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS HAD INTERMITTENT RESPONSE AND REQUIRED MULTIPLE PRESSES TO EVOKE A RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.
ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING THE OK, UP AND DOWN ARROW KEYPAD BUTTONS WERE UNRESPONSIVE. NO OTHER INFORMATION IS AVAILABLE AT THIS TIME. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |