FDA Adverse Event Injury Summary report: N

M2A-MAGNUM 42-50MM TPR INSRT-6

MDR report key: 2830197 · Received November 13, 2012

Report

Report Number
0001825034-2012-02405
Event Type
Injury
Date Received
November 13, 2012
Date of Event
January 19, 2012
Report Date
October 18, 2012
Manufacturer
BIOMET ORTHOPEDICS
Product Code
KWA
PMA / PMN Number
PK042037
Removal / Correction Number
N/A
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MT, US
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER WARNINGS, NUMBER 1 STATES, "MATERIAL SENSITIVITY REACTIONS." AND NUMBER 14, "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2012-02404 / 02406). THE REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Additional Manufacturer Narrative · 1

THIS FOLLOW-UP REPORT IS BEING FILED TO RELAY THE REVISION PROCEDURE INFORMATION AND BLOOD TEST RESULTS, WHICH WERE UNKNOWN AT THE TIME OF THE INITIAL MEDWATCH. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THIS REPORT IS NUMBER 2 OF 3 MDRS FILED FOR THE SAME EVENT (REFERENCE 1825034-2012-02404 / 02406).

Description of Event or Problem · 1

LEGAL COUNSEL FOR THE PATIENT ALLEGES THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2009. A SUBSEQUENT REVISION WAS PERFORMED (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, INCREASED METAL ION LEVELS AND METALLOSIS. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF'S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Description of Event or Problem · 1

LEGAL COUNSEL FOR THE PATIENT ALLEGES THAT PATIENT UNDERWENT RIGHT TOTAL HIP ARTHROPLASTY ON (B)(6) 2009. A SUBSEQUENT REVISION WAS PERFORMED (B)(6) 2012 DUE TO PATIENT ALLEGATIONS OF PAIN, DISCOMFORT, INCREASED METAL ION LEVELS AND METALLOSIS. ADDITIONAL INFORMATION RECEIVED IN PATIENT MEDICAL RECORDS INDICATES THAT THE REVISION PROCEDURE THAT TOOK PLACE ON (B)(6) 2012 WAS DUE TO ACETABULAR CUP LOOSENING AND ALVAL LESION. THE OPERATIVE NOTES CONFIRM THAT THE ACETABULAR CUP AND HEAD WERE REMOVED AND REPLACED. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT, AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M2A-MAGNUM 42-50MM TPR INSRT-6 PROSTHESIS, HIP KWA BIOMET ORTHOPEDICS N/A 906790

Patients

Seq Age Sex Outcome Treatment
1 57 YR Hospitalization| R