FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2830115 · Received November 13, 2012

Report

Report Number
1416980-2012-04938
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
October 19, 2012
Report Date
October 19, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE CUSTOMER REPORTED PROBLEM OF " ALARM F-38" WAS CONFIRMED. SERVICE DETERMINED THAT THE CAUSE WAS DUE TO THE FORCE SENSING RESISTORS BEING OUT OF SPECIFICATION. THE FORCE SENSING RESISTORS WERE REPLACED TO RESOLVE THE ISSUE. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A FAILURE CODE OF 38. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1