FLOGARD
Report
- Report Number
- 1416980-2012-04938
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Date of Event
- October 19, 2012
- Report Date
- October 19, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). DEVICE EVALUATION: THE CUSTOMER REPORTED PROBLEM OF " ALARM F-38" WAS CONFIRMED. SERVICE DETERMINED THAT THE CAUSE WAS DUE TO THE FORCE SENSING RESISTORS BEING OUT OF SPECIFICATION. THE FORCE SENSING RESISTORS WERE REPLACED TO RESOLVE THE ISSUE. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP WITH A FAILURE CODE OF 38. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.