FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 2830039 · Received November 13, 2012

Report

Report Number
1416980-2012-04921
Event Type
Malfunction
Date Received
November 13, 2012
Date of Event
July 19, 2012
Report Date
October 22, 2012
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Removal / Correction Number
1423500-01/08/10-001-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION: THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, TWO INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENTS WERE IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2012. DURING NIGHT DRAIN CYCLE 21, THE PATIENT'S ULTRAFILTRATION READING WAS 1602ML, INDICATING THE HOME PATIENT (HP) DRAINED 1602ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1