FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2830026 · Received November 13, 2012

Report

Report Number
2531779-2012-13504
Event Type
Malfunction
Date Received
November 13, 2012
Report Date
November 1, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW UP #1 SUBMITTED: 12/20/2012 DEVICE EVALUATION: THE INSULIN PUMP HAS BEEN RETURNED AND ADDITIONAL INVESTIGATION WAS PERFORMED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: ON INVESTIGATION, THE KEYPAD WAS FOUND TO BE FULLY INTACT WITHOUT PEELING OR VISIBLE DAMAGE. ON TESTING, ALL THE KEYPAD BUTTONS DEMONSTRATED INTERMITTENT UNRESPONSIVENESS. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE KEYPAD BUTTONS.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING THE UP AND DOWN ARROW KEYPAD BUTTONS HAD DELAYED RESPONSES. THE REPORTER DENIED DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP ATTACHED TO A BELT AND DID NOT CLEAN IT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 15 YR