ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2012-13495
- Event Type
- Malfunction
- Date Received
- November 13, 2012
- Report Date
- October 31, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DISCOLORED/FADED DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION. THE OWNER'S BOOKLET INSTRUCTS THE USER TO CALL ANIMAS CUSTOMER SERVICE IF THE USER SUSPECTS THAT THE PRODUCT IS DAMAGED.
ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS, ALLEGING ALL THE KEYPAD BUTTONS HAD INTERMITTENTLY DELAYED RESPONSES. THE PATIENT DENIED DAMAGE TO THE KEYPAD. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |