BABY CONTROL COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2012-00809
- Event Type
- Malfunction
- Date Received
- November 12, 2012
- Date of Event
- October 15, 2012
- Report Date
- October 15, 2012
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). SERIAL NUMBERS: (B)(4); LOT NUMBERS: 040517, 040517; DEVICE MANUFACTURER DATE: 05/17/2004, 05/17/2004. METHOD: THE COMPLAINT WARMER HEADS WERE NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. VISUAL INSPECTION OF PHOTOGRAPHS PROVIDED BY THE CUSTOMER WAS PERFORMED. RESULTS: THE PHOTOGRAPH INSPECTION REVEALED CRACKS ON THE DOME PORTION OF THE WARMER HEAD. CRAZING AND FLAKING OF THE PLASTIC SHELL WAS APPARENT AND CRACKS COULD BE SEEN IN THE MIDDLE OF THE SIDES OF THE UPPER CASE OF THE WARMER HEAD. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 040517. CONCLUSION: IT IS LIKELY THAT THE CRACKING AND CRAZING OBSERVED ON THE WARMER HEAD ASSEMBLY ARE RELATED TO A KNOWN PROBLEM OF INCOMPATIBLE CLEANING SOLUTIONS REACTING WITH THE POLYCARBONATE PLASTIC AND ACRYLIC COMPONENTS OF THE INFANT WARMER. WHEN THE WARMER HEAD BEGINS TO WARM UP DURING USE, A CHEMICAL REACTION WITH THE INCOMPATIBLE CLEANING SOLUTION RESULTS IN GRADUAL CRACKING AND CRAZING OF THE WARMER HEAD. THE INFANT WARMER SERVICE MANUAL FOR THE AFFECTED UNIT FEATURES A DIAGRAM OF THE INFANT WARMER WHICH HIGHLIGHTS POLYCARBONATE AND ACRYLIC COMPONENTS AND STATES THE FOLLOWING: "CAUTION DO NOT CLEAN THE HIGHLIGHTED PLASTIC SURFACES WITH PROPRIETARY CLEANING PRODUCTS CONTAINING EITHER HYDROXIDES, HYPOCHLORITES, PEROXIDES, GLUTERALDEHYDE OR CLEANING PRODUCTS WITH A GREATER THAN 30% ALCOHOL BASE"; "CAUTION THE CHEMICALS USED IN THESE PROPRIETARY CLEANING PRODUCTS MAY LEAD TO DISCOLOURATION, CRAZING AND EVENTUAL CRACKING OF THE HIGHLIGHTED PLASTIC SURFACES." AS PART OF CONTINUOUS IMPROVEMENT, WE HAVE SINCE REVISED OUR CLEANING INSTRUCTIONS TO RECOMMEND SPECIFIC PROPRIETARY CLEANING WIPES.
(B)(4). SERIAL NUMBERS: (B)(4); LOT NUMBERS: 040517, 040517; DEVICE MANUFACTURER DATE: 05/17/2004, 05/17/2004. THE COMPLAINT DEVICES ARE NOT EXPECTED TO BE RETURNED TO THE MANUFACTURER. AN EVALUATION BASED ON THE INFORMATION RECEIVED IS ANTICIPATED. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
A HOSPITAL IN (B)(6) REPORTED THAT THE UPPER HEATER HEAD CASE OF TWO IW930 BABY CONTROL COSYCOT INFANT WARMERS WERE CRACKED. THIS WAS REPORTED PRIOR TO PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT THE UPPER HEATER HEAD CASE OF TWO IW930 BABY CONTROL COSYCOT INFANT WARMERS WERE CRACKED. THIS WAS REPORTED PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BABY CONTROL COSYCOT INFANT WARMER | FMT | FMT | FISHER & PAYKEL HEALTHCARE LTD | IW930 | SEE H.10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |