FDA Adverse Event Malfunction Summary report: N

IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM

MDR report key: 2828157 · Received November 12, 2012

Report

Report Number
2050012-2012-01821
Event Type
Malfunction
Date Received
November 12, 2012
Date of Event
October 17, 2012
Report Date
October 17, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
JQX
PMA / PMN Number
K962294
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 20128, A FIELD SERVICE ENGINEER (FSE) WAS ON SITE AND FOUND A LOOSE CONNECTION BETWEEN THE WASH BUFFER SUPPLY AND THE SYRINGE SELECTOR VALVE. THE FSE SECURED THE FITTING INTERNAL PROBE WASH ASSEMBLY. THE FAILURE MODE OF LEAK WAS THE LOOSE FITTING INTERNAL PROVE WASH ASSEMBLY. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN UNCONTAINED FLUID LEAK UNDERNEATH THE IMMAGE 800 IMMUNOCHEMISTRY SYSTEMS, WHICH LEAKED ONTO THE COUNTER TOP. PER CUSTOMER THIS IS A NEW INSTRUMENT AND THEY HAVE NOT RUN PATIENT SAMPLES ONLY QUALITY CONTROLS ON IT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LAB COAT, EYE PROTECTION, AND GLOVES. THE CUSTOMER DID NOT COME IN CONTACT WITH THE FLUID AND DID NOT SEEK MEDICAL ATTENTION. NO INJURY OR EXPOSURE (SPRAYED OR SPLASHED) TO MUCOUS MEMBRANES OR OPEN WOUNDS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE® 800 IMMUNOCHEMISTRY SYSTEM NEPHELOMETER, FOR CLINICAL USE JQX BECKMAN COULTER, INC. N/A