FDA Adverse Event Injury Summary report: N

IMPLANT PDS3000 6PK PILLAR PALATAL EN

MDR report key: 2823917 · Received October 29, 2012

Report

Report Number
1045254-2012-00631
Event Type
Injury
Date Received
October 29, 2012
Date of Event
November 1, 2010
Report Date
November 18, 2010
Manufacturer
MEDTRONIC XOMED, INC.
Product Code
LRK
PMA / PMN Number
K040417
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS DISCARDED / NOT RETURNED TO THE MFR FOR EVAL. THE DEVICE WAS MANUFACTURED BY RESTORE MEDICAL, WHICH WAS SUBSEQUENTLY PURCHASED BY MEDTRONIC XOMED - THE MANUFACTURE DATE OF THIS DEVICE IS NOT AVAILABLE. THE IMPLANT IS INTENDED FOR USE IN STIFFENING THE SOFT PALATE TISSUE, WHICH MAY REDUCE THE SEVERITY OF SNORING IN SOME INDIVIDUALS, AND FOR THE REDUCTION OF THE INCIDENCE OF AIRWAY OBSTRUCTIONS IN PTS SUFFERING FROM MILD TO MODERATE OBSTRUCTIVE SLEEP APNEA (OSA). THE SYSTEM CONSISTS OF A DELIVERY TOOL AND AN IMPLANT. THE DELIVERY TOOL COMES PRELOADED WITH THE IMPLANT. (B)(4). THE IMPLANT IS APPROX 0.7 INCHES (18MM) IN LENGTH AND HAS AN APPROXIMATE OUTER DIAMETER OF 0.08 INCHES (2 MM). THE DELIVERY TOOL CONSISTS OF A HANDLE AND 14-GAUGE NEEDLE. THE NEEDLE IS INSERTED INTO THE SOFT PALATE; THE IMPLANT IS DEPLOYED BY ADVANCING THE SLIDER; AND THE DELIVERY TOOL IS REMOVED. THE DELIVERY TOOL IS DISPOSABLE. USE OF THE IMPLANT INVOLVES POTENTIAL RISKS NORMALLY ASSOCIATED WITH THE USE OF ANY IMPLANTED DEVICE, INCLUDING, BUT NOT LIMITED TO, IMPLANT MIGRATION AND PARTIAL / FULL EXTRUSION OF THE IMPLANT. THE RELATIONSHIP OF THE DEVICE TO THE REPORTED INCIDENT IS UNCLEAR. THE PRODUCT WAS DISCARDED BY THE CUSTOMER AND NOT RETURNED TO THE MFR; THEREFORE NO PRODUCT ANALYSIS IS AVAILABLE. WITHOUT RETURN OF THE DEVICE, IT CANNOT BE DETERMINED WHETHER OR NOT IT FAILED TO MEET SPEC. NO APPLICABLE IMAGING FILMS OR MEDICAL RECORDS WERE RECEIVED. NO PT INFO/IDENTIFIER WAS SUBMITTED WITH THE ORIGINAL REPORT, WITHOUT WHICH IT IS NOT POSSIBLE TO F/U WITH THE CLINICIAN. THEREFORE, THE AVAILABLE INFO IS INCONCLUSIVE AS TO THE CAUSE OF THE REPORTED PRODUCT PROBLEM. INFO RECEIVED REASONABLY SUGGESTS SERIOUS INJURY, OR MEDICAL INTERVENTION WAS REQUIRED TO PRECLUDE SERIOUS INJURY, THUS WE ARE FILING THIS REPORT AS AN ADVERSE EVENT AND PRODUCT PROBLEM.

Description of Event or Problem · 1

THIS REPORT IS PROVIDED AS A PART OF A RETROSPECTIVE REVIEW, AND WAS PERFORMED AS THE RESULT OF RECENT CHANGES/IMPROVEMENTS TO PRODUCT SPECIFIC CRITERIA DEVELOPED TO MAKE MEDICAL DEVICE REPORT (MDR) DECISIONS RELATED TO THE PILLAR PALATIAL IMPLANTS, PER DISCUSSION WITH OFFICE OF SURVEILLANCE AND BIOMETRICS (OSB). THIS CHANGE/IMPROVEMENT, WHICH WAS INITIATED IN JANUARY 2012, HAS RESULTED IIN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY THE MFR - THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF ANY NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE SUBMISSION CRITERIA. FIVE PROSTHESIS PALATIAL ((B)(4)) IMPLANTS WERE IMPLANTED INTO THE SOFT PALATE OF THIS PT. A FEEDBACK FORM RECEIVED FROM THE PHYSICIAN SUGGESTS ONE IMPLANT REMOVAL AND ALSO INDICATES PAIN OR DIFFICULTY SWALLOWING, AND FOREIGN BODY SENSATION APPROX 25 MONTHS AFTER IMPLANT. F/U FOUND ONE IMPLANT EXTRUDED AND WAS REMOVED; AFTER IMPLANT. F/U FOUND ONE IMPLANT EXTRUDED AND WAS REMOVED PRODUCT WAS DISCARDED - NOT RETURNED FOR EVAL. REPLACEMENT WAS REQUESTED; OTHER INFO IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMPLANT PDS3000 6PK PILLAR PALATAL EN LRK - DEVICE, ANTI-SNORING LRK MEDTRONIC XOMED, INC. PDS3000 8208

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention