IMPLANT PDS3000M 6PK PILLAR PALATAL
Report
- Report Number
- 1045254-2012-00651
- Event Type
- Injury
- Date Received
- October 29, 2012
- Date of Event
- December 5, 2011
- Report Date
- December 14, 2011
- Manufacturer
- XOMED MFG JACKSONVILLE
- Product Code
- LRK
- PMA / PMN Number
- K040417
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE DEVICE WAS DISCARDED / NOT RETURNED TO THE MANUFACTURER FOR EVALUATION. HOWEVER, A REVIEW OF DEVICE HISTORY RECORDS (DHR) FOUND NO NON-CONFORMANCES IN THE PRODUCTION OF THIS LOT. METHOD: REVIEW OF DEVICE HISTORY RECORDS PERFORMED. THE IMPLANT IS INTENDED FOR USE IN STIFFENING THE SOFT PALATE TISSUE, WHICH MAY REDUCE THE SEVERITY OF SNORING IN SOME INDIVIDUALS, AND FOR THE REDUCTION OF THE INCIDENCE OF AIRWAY OBSTRUCTIONS IN PTS SUFFERING FROM MILD TO MODERATE OBSTRUCTIVE SLEEP APNEA (OSA). THE SYSTEM CONSISTS OF A DELIVERY TOOL AND AN IMPLANT. THE DELIVERY TOOL COMES PRELOADED WITH THE IMPLANT. (B)(4). THE IMPLANT IS APPROXIMATELY 0.7 INCHES (18MM) IN LENGTH AND HAS AN APPROXIMATE OUTER DIAMETER OF 0.08 INCHES (2MM). THE DELIVERY TOOL CONSISTS OF A HANDLE AND 14- GAUGE NEEDLE. THE NEEDLE IS INSERTED INTO THE SOFT PALATE; THE IMPLANT IS DEPLOYED BY ADVANCING THE SLIDER; AND THE DELIVERY TOOL IS REMOVED. THE DELIVERY TOOL IS DISPOSABLE. USE OF THE IMPLANT INVOLVES POTENTIAL RISKS NORMALLY ASSOCIATED WITH THE USE OF ANY IMPLANTED DEVICE, INCLUDING, BUT NOT LIMITED TO, IMPLANT MIGRATION AND PARTIAL / FULL EXTRUSION OF THE IMPLANT. THE RELATIONSHIP OF THE DEVICE TO THE REPORTED INCIDENT IS UNCLEAR. THE PRODUCT WAS DISCARDED BY THE CUSTOMER AND NOT RETURNED TO THE MANUFACTURER; THEREFORE NO PRODUCT ANALYSIS IS AVAILABLE. HOWEVER, A REVIEW OF DEVICE HISTORY RECORDS (DHR) FOUND NO NON-CONFORMANCES IN THE PRODUCTION OF THIS LOT. WITHOUT RETURN OF THE DEVICE, IT CANNOT BE DETERMINED WHETHER OR NOT IT FAILED TO MEET SPECIFICATION. NO APPLICABLE IMAGING FILMS OR MEDICAL RECORDS WERE RECEIVED. NO PT INFORMATION/IDENTIFIER WAS SUBMITTED WITH THE ORIGINAL REPORT, WITHOUT WHICH IT IS NOT POSSIBLE TO FOLLOW-UP WITH THE CLINICIAN. THEREFORE, THE AVAILABLE INFORMATION IS INCONCLUSIVE AS TO THE CAUSE OF THE REPORTED PRODUCT PROBLEM. INFORMATION RECEIVED REASONABLY SUGGESTS SERIOUS INJURY, OR MEDICAL INTERVENTION TO PRECLUDE SERIOUS INJURY, THUS WE ARE FILING THIS REPORT AS AN ADVERSE EVENT AND PRODUCT PROBLEM.
THIS REPORT IS PROVIDED AS A PART OF A RETROSPECTIVE REVIEW RELATED TO MEDICAL DEVICE REPORT (MDR) DECISIONS REGARDING PILLAR PALATIAL IMPLANTS. THIS CHANGE/IMPROVEMENT, WHICH WAS INITIATED IN JANUARY 2012, HAS RESULTED IN AN OVERALL INCREASE IN THE NUMBER OF MDRS FILED BY THE MANUFACTURER - THIS INCREASE IS NOT THE RESULT OF THE DISCOVERY OF ANY NEW PRODUCT PROBLEMS, BUT RATHER THE RESULT OF A BROADER INTERPRETATION AND APPLICATION OF THE SUBMISSION CRITERIA. FOUR PROSTHESIS PALATAL ((B)(4)) IMPLANTS WERE IMPLANTED INTO THE SOFT PALATE OF THIS PT. A FEEDBACK FORM RECEIVED FROM THE PHYSICIAN INDICATES THAT THE PT PRESENTED APPROXIMATELY THREE MONTHS POST PROCEDURE WITH ONE IMPLANT PARTIALLY POKING THROUGH THE MUCOSA; IT WAS TRIMMED. ONE MONTH LATER, THE IMPLANT WAS AGAIN EXTRUDING AND TRIMMED, "MAKING IT NON-FUNCTIONAL." AT THAT TIME A SEPARATE IMPLANT HAD EXTRUDED AND WAS COMPLETELY REMOVED. THE PRODUCT WAS DISCARDED - NOT RETURNED FOR EVALUATION. FOLLOW-UP FOUND "THE PT IS DOING FINE." NO OTHER INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMPLANT PDS3000M 6PK PILLAR PALATAL | LRK - DEVICE, ANTI-SNORING | LRK | XOMED MFG JACKSONVILLE | PDS3000M | 73034200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Required Intervention |