FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2823772 · Received October 10, 2012

Report

Report Number
1824206-2012-06354
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
April 24, 2012
Report Date
April 24, 2012
Manufacturer
HILL-ROM INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN ISOLATED THE ISSUE TO HYDRAULIC MANIFOLD. HE REPLACED THE HYDRAULIC MANIFOLD TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED IS SLOWLY DRIFTING INTO TRENDELENBURG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM INC. 1900

Patients

Seq Age Sex Outcome Treatment
1