FDA Adverse Event
Malfunction
Summary report: N
TOTALCARE BED
MDR report key: 2823750
·
Received October 10, 2012
Report
- Report Number
- 1824206-2012-06343
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- September 24, 2012
- Report Date
- September 24, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE HEAD SECTION WAS DRIFTING DOWN 5 DEGREES OVER 20 MINUTES. HE ISOLATED THE ISSUE TO A STICKING HEAD DOWN VALVE. HE REPLACED THE HEAD DOWN VALVE TO REPAIR THE BED.
Description of Event or Problem · 1
THE COMPLAINANT STATED THAT THE HEAD SECTION IS DRIFTING DOWN AFTER THE FUNCTION SWITCH IS RELEASED. IT TAKES A FEW MINUTES TO DRIFT DOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TOTALCARE BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 1900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |