FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2823750 · Received October 10, 2012

Report

Report Number
1824206-2012-06343
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 24, 2012
Report Date
September 24, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE HEAD SECTION WAS DRIFTING DOWN 5 DEGREES OVER 20 MINUTES. HE ISOLATED THE ISSUE TO A STICKING HEAD DOWN VALVE. HE REPLACED THE HEAD DOWN VALVE TO REPAIR THE BED.

Description of Event or Problem · 1

THE COMPLAINANT STATED THAT THE HEAD SECTION IS DRIFTING DOWN AFTER THE FUNCTION SWITCH IS RELEASED. IT TAKES A FEW MINUTES TO DRIFT DOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1