FDA Adverse Event
Malfunction
Summary report: N
CENTRA BED
MDR report key: 2823726
·
Received October 10, 2012
Report
- Report Number
- 1824206-2012-06322
- Event Type
- Malfunction
- Date Received
- October 10, 2012
- Date of Event
- September 20, 2012
- Report Date
- September 20, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECH REPLACED THE STIFFENER PLATE AND INSTALLED NEW BRAKE BEARINGS IN THE BRAKE MECHANISM BLOCK AND REPLACED THE BRAKE CASTER TO RESOLVE THE ISSUE.
Description of Event or Problem · 1
THE TECH ALLEGED THAT THE BRAKES ARE NOT HOLDING. NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CENTRA BED | AC-POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 850 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |