FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 2823726 · Received October 10, 2012

Report

Report Number
1824206-2012-06322
Event Type
Malfunction
Date Received
October 10, 2012
Date of Event
September 20, 2012
Report Date
September 20, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE STIFFENER PLATE AND INSTALLED NEW BRAKE BEARINGS IN THE BRAKE MECHANISM BLOCK AND REPLACED THE BRAKE CASTER TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECH ALLEGED THAT THE BRAKES ARE NOT HOLDING. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850

Patients

Seq Age Sex Outcome Treatment
1