BIPOLAR CUTTING LOOP ELECTRODE - DEVICE #3
Report
- Report Number
- 1418479-2010-00024
- Event Type
- Malfunction
- Date Received
- September 3, 2010
- Date of Event
- August 9, 2010
- Report Date
- September 3, 2010
- Manufacturer
- RICHARD WOLF MEDICAL INSTRUMENTS
- Product Code
- KNS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
EVAL SUMMARY: (B)(4) - BIPOLAR COAGULATING ELECTRODE. VISUAL INSPECTION CONFIRMED THE WIRE LOOP BREAKAGE. HEAVY BURN MARKS WERE EVIDENT ON BOTH WIRE LOOPS. BLUE INSULATION WAS MELTED. USER FACILITY CONFIRMED THAT THIS PROCEDURE WAS EITHER THE FIRST OR SECOND ONE THAT THE SURGEON HAD PERFORMED WITH THE RICHARD WOLF ELECTRODES. WE WERE INFORMED THAT THE SURGEON, CIRCULATING NURSE AND A TECH WAS IN THE ROOM FOR THE PROCEDURE. THERE WAS NO RICHARD WOLF SALES REP THERE FOR THIS PROCEDURE. THE FACILITY DID NOT PROVIDE A LOT NUMBER FOR THE ELECTRODES. CAUSE OF EVENT: A HIGH ELECTROSURGICAL SETTING WAS USED. IN ADDITION, THE SURGEON MUST BE MINDFUL OF THE SPEED OF RESECTION. TECHNIQUE SERVICES AS THE CAUSE OF THE DEVICE FAILURE. WE HAVE ADDRESSED THIS VIA LETTER TO THE FACILITY. IN ADDITION, OUR CURRENT SALES REP IS CLOSELY INVOLVED WITH THE FACILITY AND WILL ASSIST WITH THEIR NEEDS.
IT WAS REPORTED THAT DURING A TRANSURETHRAL RESECTION OF THE PROSTATE PROCEDURE, FOUR WIRE LOOPS ELECTRODES BROKE IN THE PT AND THE PIECES WERE RETRIEVED FROM THE PT. THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BIPOLAR CUTTING LOOP ELECTRODE - DEVICE #3 | BIPOLAR CUTTING LOOP ELECTRODE | KNS | RICHARD WOLF MEDICAL INSTRUMENTS | 4623.0223 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |