LT 270-54, EASYPUMP, 270ML, 5ML/HR
Report
- Report Number
- 2026095-2012-00262
- Event Type
- Other
- Date Received
- November 2, 2012
- Date of Event
- April 30, 2012
- Report Date
- October 10, 2012
- Manufacturer
- I-FLOW LLC
- Product Code
- MEB
- PMA / PMN Number
- K052117
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
METHOD: THE ACTUAL DEVICE INVOLVED IN THE INCIDENT WAS RECEIVED FOR EVAL AND INVESTIGATION. A REVIEW OF THE DEVICE HISTORY RECORD WAS CONDUCTED FOR THE LOT NUMBER REPORTED. THE DEVICE LOT MET MFG SPECIFICATIONS PRIOR TO RELEASE. A VISUAL EXAM WAS PERFORMED. THE PUMP WAS RECEIVED EMPTY. AN ATTEMPT TO FILL THE PUMP TO NOMINAL VALUE OF 270ML WAS MADE. RESULTS: THE PUMP LEAKED IMMEDIATELY WHEN THE FILLING PROCESS BEGAN. CONCLUSIONS: THE COMPLAINT WAS CONFIRMED. THE CAUSE OF THE LEAK WAS ATTRIBUTED TO A BROKEN O-RING. PER PROCEDURE PR-00598-CORRECTIVE AND PREVENTATIVE ACTION- THIS INCIDENT OF A BROKEN O-RING IS BEING INVESTIGATED UNDER CAPA (B)(4). LEAKING DURING THE FILLING PROCESS DOES NOT MEET THE CRITERIA FOR A REPORTABLE EVENT. (B)(4). (2026095-2012-00262).
DRUG/DILUENT: 5FU. FILL VOLUME: UNK. FLOW RATE: 5.0ML/HR. PROCEDURE: NA. CATHPLACE: NA. PT CONTACT: NO. THE PUMP LEAKED DURING THE FILLING PROCEDURE DURING THE INJECTION OF THE 3RD SYRINGE; LOCATION OF LEAK UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LT 270-54, EASYPUMP, 270ML, 5ML/HR | ELASTOMERIC PUMP | MEB | I-FLOW LLC | 04438043 | 192387 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |