FDA Adverse Event Malfunction Summary report: N

COULTER® LH 500 HEMATOLOGY ANALYZER

MDR report key: 2823430 · Received November 8, 2012

Report

Report Number
1061932-2012-02676
Event Type
Malfunction
Date Received
November 8, 2012
Date of Event
October 8, 2012
Report Date
October 11, 2012
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND NOTED NO HARDWARE ISSUES. THE INSTRUMENT WAS PERFORMING WITHIN THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS. THE LABORATORY PERFORMS DAILY MODE-TO-MODE VERIFICATION PROCEDURE, WHICH PASSED WITH ACCEPTABLE COEFFICIENT OF VARIATION (CV) LIMITS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ERRONEOUS ANALYTE RESULTS THROUGH MANUAL MODE, FOR TWO PATIENTS, INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. SAMPLE #1 INDICATES THE INITIAL MANUAL MODE RESULTS RECOVERED LOWER FOR WHITE BLOOD CELL (WBC), HEMOGLOBIN (HGB), MEAN CELL HEMOGLOBIN (MCH) AND MEAN CORPUSCULAR HEMOGLOBIN CONCENTRATION (MCHC) AND SLIGHTLY HIGHER FOR RED BLOOD CELL (RBC) RESULTS WITHOUT INSTRUMENT GENERATED FLAGS. THE HEMOGLOBIN AND HEMATOCRIT (H&H) RATIO DID NOT CORRELATE ON THE INITIAL ANALYSIS. SAMPLE #2 WAS ANALYZED FIVE TIMES IN THE MANUAL MODE. THE H&H RATIO DID NOT CORRELATE ON THE REANALYSIS, WHICH RECOVERED LOW WBC AND HIGH RBC AND PLT WITHOUT INSTRUMENT GENERATED MESSAGES EXCEPT FOR RETEST #1 AND #3. SUBSEQUENT ANALYSIS OF THE PATIENT SAMPLE, ON THE SAME INSTRUMENT, RECOVERED HIGHER RESULTS WHICH WERE CONSIDERED TO BE CORRECT. THE ERRONEOUS RESULTS WERE RELEASED OUT OF THE LABORATORY, AND AMENDED REPORTS WERE ISSUED. THERE WAS NO IMPACT TO PATIENT OUTCOME. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COULTER® LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1