DOUBLE FENESTRATED GRASPER INSTRUMENT
Report
- Report Number
- 2955842-2012-00835
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Date of Event
- September 20, 2012
- Report Date
- October 9, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT JERKING MOVEMENT IS CONFIRMED. PITCH CABLE WAS FOUND LOOSE AT DISTAL CLEVIS HUB. BOTH CABLE CRIMPS REMAINED IN CLEVIS AND NO DAMAGE WAS FOUND ON CABLE. UPON HOUSING REMOVAL FOUND PITCH CABLE LOOSE AT CLAMPING PULLEY, BUT NO LOOSE CLAMPING PULLEY SCREW WAS FOUND. ADDITIONAL FINDING: SCRATCHES ON MAINTUBE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SUGGESTING THE MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.
IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED EXPERIENCING A JERKING MOVEMENT AND NOTICED A CABLE WAS OFF OF ITS PULLEY. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DOUBLE FENESTRATED GRASPER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420189-07 | M10120308 543 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |