FDA Adverse Event Malfunction Summary report: N

DOUBLE FENESTRATED GRASPER INSTRUMENT

MDR report key: 2823389 · Received November 7, 2012

Report

Report Number
2955842-2012-00835
Event Type
Malfunction
Date Received
November 7, 2012
Date of Event
September 20, 2012
Report Date
October 9, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. COMPLAINT JERKING MOVEMENT IS CONFIRMED. PITCH CABLE WAS FOUND LOOSE AT DISTAL CLEVIS HUB. BOTH CABLE CRIMPS REMAINED IN CLEVIS AND NO DAMAGE WAS FOUND ON CABLE. UPON HOUSING REMOVAL FOUND PITCH CABLE LOOSE AT CLAMPING PULLEY, BUT NO LOOSE CLAMPING PULLEY SCREW WAS FOUND. ADDITIONAL FINDING: SCRATCHES ON MAINTUBE. DISTAL END OF MAIN TUBE HAS VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. SUGGESTING THE MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGICAL STAFF REPORTED EXPERIENCING A JERKING MOVEMENT AND NOTICED A CABLE WAS OFF OF ITS PULLEY. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DOUBLE FENESTRATED GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420189-07 M10120308 543

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES