MEGASUTURECUT NEEDLE DRIVER INSTRUMENT
Report
- Report Number
- 2955842-2012-00813
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Date of Event
- October 8, 2012
- Report Date
- October 8, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED THAT POST OP A SUCCESSFUL DA VINCI SI PARTIAL NEPHRECTOMY PROCEDURE, THE WIRES AT THE TIP OF THE MEGASUTURECUT NEEDLE DRIVER INSTRUMENT WERE ALL OUT. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MEGASUTURECUT NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420309-01 | M10120516 333 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 51 YR | DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES |