FDA Adverse Event Malfunction Summary report: N

MEGASUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 2823040 · Received November 7, 2012

Report

Report Number
2955842-2012-00813
Event Type
Malfunction
Date Received
November 7, 2012
Date of Event
October 8, 2012
Report Date
October 8, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE IS BROKEN AT THE DISTAL IDLERS. THE IDLER PULLEY SPINS FREELY AND WAS NOT DAMAGED. THE CABLE SEGMENT STICKS OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT POST OP A SUCCESSFUL DA VINCI SI PARTIAL NEPHRECTOMY PROCEDURE, THE WIRES AT THE TIP OF THE MEGASUTURECUT NEEDLE DRIVER INSTRUMENT WERE ALL OUT. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEGASUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120516 333

Patients

Seq Age Sex Outcome Treatment
1 51 YR DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES