FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2823026
·
Received November 7, 2012
Report
- Report Number
- 2531779-2012-13247
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Report Date
- October 11, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4)2012 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR |