FDA Adverse Event Malfunction Summary report: N

XIA 3 TITANIUM TORQUE WRENCH SHORT

MDR report key: 2822993 · Received October 24, 2012

Report

Report Number
9617544-2012-00458
Event Type
Malfunction
Date Received
October 24, 2012
Date of Event
September 26, 2012
Report Date
September 26, 2012
Manufacturer
STRYKER SPINE BORDEAUX
Product Code
LXH
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.

Description of Event or Problem · 1

DR. (B)(6) WAS DOING A T3 TO L3 SCOLIOSIS CASE. HE FINAL TIGHTENED ABOUT 8 BLOCKERS AND ON THE PATIENTS LEFT SIDE AND AROUND THE T8 LEVEL AND WE HEARD A SNAP. THE HEX TIP BROKE AND IS NOW STUCK ON THE BLOCKER. HE USED A DRILL AND MADE A SMALL HOLE IN THE BROKEN TIP AND WAS ABLE TO EXTRACT IT THEN WITH A SNAP. HE DID ONE MORE WASH AND CONTINUED TO FINAL TIGHTEN WITH THE OTHER TORQUE WRENCH THAT WE HAD. THIS ENTIRE PROCESS ADDED ABOUT 3 TO 5 MINUTES TO THE CASE. THE FINAL X-RAY WAS DETERMINED BY THE RADIOLOGIST THAT NO PIECES OR DEBRIS BESIDES THE IMPLANTS WERE INSIDE THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 XIA 3 TITANIUM TORQUE WRENCH SHORT INSTRUMENT LXH STRYKER SPINE BORDEAUX NA 11A957

Patients

Seq Age Sex Outcome Treatment
1 UNK