XIA 3 TITANIUM TORQUE WRENCH SHORT
Report
- Report Number
- 9617544-2012-00458
- Event Type
- Malfunction
- Date Received
- October 24, 2012
- Date of Event
- September 26, 2012
- Report Date
- September 26, 2012
- Manufacturer
- STRYKER SPINE BORDEAUX
- Product Code
- LXH
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- RI, US
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF MADE AVAILABLE WILL BE REPORTED IN A SUPPLEMENTAL. METHOD, RESULT, AND CONCLUSION CODES WILL BE MADE AVAILABLE FOLLOWING AN ENGINEERING EVALUATION.
DR. (B)(6) WAS DOING A T3 TO L3 SCOLIOSIS CASE. HE FINAL TIGHTENED ABOUT 8 BLOCKERS AND ON THE PATIENTS LEFT SIDE AND AROUND THE T8 LEVEL AND WE HEARD A SNAP. THE HEX TIP BROKE AND IS NOW STUCK ON THE BLOCKER. HE USED A DRILL AND MADE A SMALL HOLE IN THE BROKEN TIP AND WAS ABLE TO EXTRACT IT THEN WITH A SNAP. HE DID ONE MORE WASH AND CONTINUED TO FINAL TIGHTEN WITH THE OTHER TORQUE WRENCH THAT WE HAD. THIS ENTIRE PROCESS ADDED ABOUT 3 TO 5 MINUTES TO THE CASE. THE FINAL X-RAY WAS DETERMINED BY THE RADIOLOGIST THAT NO PIECES OR DEBRIS BESIDES THE IMPLANTS WERE INSIDE THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | XIA 3 TITANIUM TORQUE WRENCH SHORT | INSTRUMENT | LXH | STRYKER SPINE BORDEAUX | NA | 11A957 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |