FDA Adverse Event Malfunction Summary report: N

DUAL COOLER/HEATER

MDR report key: 2822754 · Received November 2, 2012

Report

Report Number
1828100-2012-01419
Event Type
Malfunction
Date Received
November 2, 2012
Date of Event
October 9, 2012
Report Date
October 9, 2012
Manufacturer
TERUMO CARDIOVASCULAR SYSTEM CORP
Product Code
DWC
PMA / PMN Number
PREAMEND
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REP (FSR). THE FSR REPAIRED THE LEAK. THE UNIT WAS TESTED TO MANUFACTURER'S SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. ADDITIONAL INFO BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE FIELD SERVICE REP (FSR) REPORTED THAT DURING PREVENTATIVE MAINTENANCE (PM) OF THE DEVICE, THE FSR DISCOVERED A LEAK FROM PVC PIPING AND TUBING NEAR THE THERMOSTAT. SINCE THE EVENT OCCURRED DURING PREVENTATIVE MAINTENANCE, THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DUAL COOLER/HEATER COOLING & HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEM CORP 11160

Patients

Seq Age Sex Outcome Treatment
1