FDA Adverse Event Injury Summary report: N

VA-LCP CONDYLAR PLATE 4.5/5.0 R 16HO L33

MDR report key: 2822670 · Received November 7, 2012

Report

Report Number
8030965-2012-01206
Event Type
Injury
Date Received
November 7, 2012
Report Date
October 9, 2012
Manufacturer
SYNTHES GMBH
Product Code
HRS
PMA / PMN Number
K110354
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION COULD NOT BE COMPLETED, NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM. THE INVESTIGATION IS ONGOING. NOTE: BLANK FIELDS ON THIS FORM INDICATE INFORMATION THAT IS UNKNOWN, UNAVAILABLE OR UNCHANGED. PLACEHOLDER.

Additional Manufacturer Narrative · 1

(B)(4): MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND. (B)(4): PLACEHOLDER.

Additional Manufacturer Narrative · 1

DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. ADD'L PRODUCT CODES: JDP, HWC. THE EXAMINATION OF THE RAW MATERIAL INSPECTION SHEETS AND THE MANUFACTURING PAPERS SHOWS THAT THE CHEMICAL COMPOSITION, MICROSTRUCTURE AND MECHANICAL PROPERTIES ARE IN COMPLIANCE WITH THE INTERNATIONAL STANDARDS ISO (B)(4) AND (B)(4). THE DIMENSIONS WERE FOUND TO BE IN ACCORDANCE WITH THE TECHNICAL DRAWING OF THE PRODUCER AND AO ASIF SPECIFICATIONS. WE ASSUME THAT AXIAL AND DOMINANT DYNAMIC BENDING LOADS LET TO THE FATIGUE OF THE PLATE. THE PLATE HAD TO ABSORB AND NEUTRALIZE FORCES THAT MAY HAVE BEEN GREATER THAN THE TOLERABLE LOAD. POSTOPERATIVE ACTIVITIES OF THE PATIENT MAY HAVE PLAYED A CERTAIN ROLE TOO. WE FOUND NO EVIDENCE OF MATERIAL OR MANUFACTURING DEFECTS.

Description of Event or Problem · 1

A DEVICE REPORT FROM (B)(6) INDICATED A HOSPITAL IN (B)(6) REPORTED: PATIENT WAS IMPLANTED WITH VA-LCP CONDYLAR PLATE ON (B)(6) 2012. APPROXIMATELY SIX WEEKS LATER, (B)(6) 2012, THE PLATE WAS FOUND BROKEN. PLATE WAS REMOVED, DATE UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VA-LCP CONDYLAR PLATE 4.5/5.0 R 16HO L33 VA-LCP CONDYLAR PLATE HRS SYNTHES GMBH 7677704

Patients

Seq Age Sex Outcome Treatment
1 93 YR Required Intervention PLATE, SCREWS