FDA Adverse Event Malfunction Summary report: N

ARCHITECT C4000 ANALYZER

MDR report key: 2822650 · Received November 7, 2012

Report

Report Number
1628664-2012-00469
Event Type
Malfunction
Date Received
November 7, 2012
Report Date
November 2, 2012
Manufacturer
ABBOTT MANUFACTURING INC
Product Code
JJE
PMA / PMN Number
K980367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, AND A REVIEW OF LABELING. THE CUSTOMER STATED THE FRONT COVER DOES NOT STAY UP, BUT INSTEAD SLOWLY DESCENDS. NO INJURIES HAVE BEEN REPORTED DUE TO THE FRONT COVER NOT REMAINING UP. THE FIELD SERVICE ENGINEER (FSE) ADJUSTED THE RETENTION SPRING OF THE COVER LID AND RESOLVED THE ISSUE. THE INSTRUMENT IS PERFORMING AS INTENDED AND NO FURTHER ISSUES HAVE OCCURRED. TRACKING AND TRENDING IDENTIFIED NO ADVERSE TRENDS FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. BASED ON THE AVAILABLE INFORMATION NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE FRONT COVER OF AN ARCHITECT C4000 ANALYZER FAILED TO LATCH. THERE WAS NO INJURY TO PERSONNEL REPORTED. THE TENSION IN THE COVER HINGE WAS ADJUSTED TO RESOLVE THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ARCHITECT C4000 ANALYZER JJE ABBOTT MANUFACTURING INC

Patients

Seq Age Sex Outcome Treatment
1