FDA Adverse Event Malfunction Summary report: N

MONOPOLAR CAUTERY INSTRUMENT

MDR report key: 2822608 · Received November 7, 2012

Report

Report Number
2955842-2012-00803
Event Type
Malfunction
Date Received
November 7, 2012
Date of Event
July 12, 2012
Report Date
October 8, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT THE CAUTERY CONNECTOR TIP WAS BROKEN AT THE SNAKE WRIST LOCATION. THE ELECTRICAL PRONG WAS EXPOSED AS RESULT. ENGINEERING CONCLUDED THAT DAMAGE WAS LIKELY DUE TO MISHANDLING. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING AND STERILIZATION PROCESS, THE CUSTOMER NOTED THAT THE MONOPOLAR CAUTERY INSTRUMENT TIP WAS DAMAGED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MONOPOLAR CAUTERY INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420142-04 S10120213 241

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES