CUSTOM DEFINED PRODUCT
Report
- Report Number
- 2015691-2012-18616
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Date of Event
- September 23, 2012
- Report Date
- October 18, 2012
- Manufacturer
- EDWARDS LIFESCIENCES DR
- Product Code
- DXO
- PMA / PMN Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT WAS NOT RETURNED FOR EVALUATION. A REVIEW OF THE MANUFACTURING RECORDS COULD NOT BE DONE WITHOUT A LOT NUMBER. THE DPT INSTRUCTIONS FOR USE, PROVIDE THE FOLLOWING INFORMATION AND GUIDANCE RELATED TO ABNORMAL PRESSURE READINGS: PRESSURE READINGS CAN CHANGE QUICKLY AND DRAMATICALLY BECAUSE OF LOSS OF PROPER CALIBRATION, LOOSE CONNECTION, OR AIR IN THE SYSTEM. WARNING: ABNORMAL PRESSURE READINGS SHOULD CORRELATE WITH THE PATIENT'S CLINICAL MANIFESTATIONS. VERIFY TRANSDUCER FUNCTION WITH A KNOWN AMOUNT OF PRESSURE BEFORE INSTITUTING THERAPY. CAUTION: SIGNIFICANT DISTORTION OF THE PRESSURE WAVEFORM OR AIR EMBOLI CAN RESULT FROM AIR BUBBLES IN THE SETUP. NOTE: POOR DYNAMIC RESPONSE CAN BE CAUSED BY AIR BUBBLES, CLOTTING, EXCESSIVE LENGTHS OF TUBING, EXCESSIVELY COMPLIANT PRESSURE TUBING, SMALL BORE TUBING, LOOSE CONNECTIONS, OR LEAKS. WITHOUT RETURN OF THE PRODUCT, IT COULD NOT BE DETERMINED IF DAMAGES OR DEFECTS WERE PRESENT ON THE DPT. IT REMAINS UNKNOWN IF ANY CLINICAL FACTORS MAY HAVE CONTRIBUTED TO THE EVENT. NO ACTIONS WILL BE TAKEN AT THIS TIME.
AS REPORTED, THE ARTERIAL BLOOD PRESSURE MEASURED ON THE DPT, GRADUALLY INCREASED TO 230/180. THE CLINICIAN DOUBTED THE ACCURACY OF THE PRESSURE MEASUREMENT, SO THEY CHECKED IT MANUALLY (CUFF PRESSURE) AND GOT A READING OF 105/56. AFTER CHANGING THE TRANSDUCER, THE ARTERIAL PRESSURE MEASURED 105/51. AS REPORTED, THE NORADRENALIN WAS RESTARTED. NO ACTUAL INJURY WAS REPORTED; HOWEVER, THE CLINICIAN CONSIDERED THAT THEY HAD GIVEN A "LIFE-THREATENING PROGNOSIS AND AMINE TREATMENT" TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CUSTOM DEFINED PRODUCT | DISPOSABLE PRESSURE TRANSDUCER | DXO | EDWARDS LIFESCIENCES DR | T100209A | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |