HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2012-04276
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Date of Event
- October 31, 2012
- Report Date
- October 31, 2012
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PATIENT
Narratives
(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) IS CONFIRMED BECAUSE THE PATIENT REPORTED HAVING A CLAMP OPEN ON AN UNUSED SUPPLY LINE DURING THERAPY, WHICH IS USE ERROR THAT CAN CAUSE SYSTEM ERROR 2240 ALARMS. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) DETERMINED THE HOME PATIENT LEFT THE BLUE CLAMP LINE OPEN. THE CAUSE WAS DETERMINED TO BE USE ERROR. A LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE PREVENTION OF THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL INFORMATION IS RECEIVED.
A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240/2367 (AIR IN LINE) DURING DWELL 3 OF 4 ON THE HOME CHOICE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) DETERMINED THAT TWO BAGS WERE CONNECTED, ONE ON THE RED CLAMP LINE AND ONE ON THE WHITE CLAMP LINE. THE TSR DETERMINED THE HOME PATIENT (HP) HAD LEFT THE BLUE CLAMP LINE OPEN AND EXPLAINED THE IMPORTANCE OF CLOSING ANY CLAMPS IN WHICH THE BAGS ARE NOT ATTACHED. THE TSR INSTRUCTED THE HP TO CLOSE ALL THE CLAMPS TO THE BAGS/PATIENT LINE/TRANSFER SET, CYCLE THE POWER OFF/ON TO GET THE SYSTEM ERROR 2367, CYCLE THE POWER OFF/ON AGAIN, PRESS GO TO START, AND AT THE LOAD THE SET, OPEN THE CASSETTE DOOR. THE TSR ADVISED THE HP TO DISPOSE OF THE CURRENT SUPPLIES AND CONTINUE WITH EITHER MANUAL BAGS OR ALL NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | HOMECHOICE |