FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2822545 · Received November 7, 2012

Report

Report Number
1416980-2012-04276
Event Type
Malfunction
Date Received
November 7, 2012
Date of Event
October 31, 2012
Report Date
October 31, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS COMPLAINT FOR A SYSTEM ERROR (SE) 2240 (AIR IN SET) IS CONFIRMED BECAUSE THE PATIENT REPORTED HAVING A CLAMP OPEN ON AN UNUSED SUPPLY LINE DURING THERAPY, WHICH IS USE ERROR THAT CAN CAUSE SYSTEM ERROR 2240 ALARMS. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) DETERMINED THE HOME PATIENT LEFT THE BLUE CLAMP LINE OPEN. THE CAUSE WAS DETERMINED TO BE USE ERROR. A LABELING REVIEW FOUND THE LABELING TO BE ADEQUATE FOR THE PREVENTION OF THE USE/USER ERROR IDENTIFIED IN THIS INCIDENT. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REPORTING A SYSTEM ERROR (SE) 2240/2367 (AIR IN LINE) DURING DWELL 3 OF 4 ON THE HOME CHOICE (HC). THE TECHNICAL SERVICE REPRESENTATIVE (TSR) DETERMINED THAT TWO BAGS WERE CONNECTED, ONE ON THE RED CLAMP LINE AND ONE ON THE WHITE CLAMP LINE. THE TSR DETERMINED THE HOME PATIENT (HP) HAD LEFT THE BLUE CLAMP LINE OPEN AND EXPLAINED THE IMPORTANCE OF CLOSING ANY CLAMPS IN WHICH THE BAGS ARE NOT ATTACHED. THE TSR INSTRUCTED THE HP TO CLOSE ALL THE CLAMPS TO THE BAGS/PATIENT LINE/TRANSFER SET, CYCLE THE POWER OFF/ON TO GET THE SYSTEM ERROR 2367, CYCLE THE POWER OFF/ON AGAIN, PRESS GO TO START, AND AT THE LOAD THE SET, OPEN THE CASSETTE DOOR. THE TSR ADVISED THE HP TO DISPOSE OF THE CURRENT SUPPLIES AND CONTINUE WITH EITHER MANUAL BAGS OR ALL NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 37 YR HOMECHOICE