FDA Adverse Event Malfunction Summary report: N

PLASMABLADE TISSUE DISSECTION DEVICE

MDR report key: 2822313 · Received October 25, 2012

Report

Report Number
3007069406-2012-00431
Event Type
Malfunction
Date Received
October 25, 2012
Report Date
August 25, 2011
Manufacturer
MEDTRONIC ADVANCED ENERGY LLC
Product Code
GEI
PMA / PMN Number
K083415
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS MDR IS BEING FILED BASED ON A RETROSPECTIVE REVIEW OF COMPLAINTS RECEIVED BY THE MANUFACTURER FOR THE TIME PERIOD (B)(4) 2010 THROUGH (B)(4) 2012. VISUAL INSPECTION OF THE RETURNED DEVICE REVEALED OF THE ADENOID TIP REVEALED THE ELECTRODE WIRE WAS BROKEN AND THE CLEAR HOUSING HAD SLIGHT MELT BACK. THE LOCATION OF THE ELECTRODE WERE BRAKE WAS AT THE CORNER OF THE HOUSING. EXCESSIVE CLEANING DURING USE CONTRIBUTED TO THE WIRE BREAKAGE. REVIEW OF THE LOT HISTORY REVEALED NO ANOMALIES. END OF REPORT.

Description of Event or Problem · 1

IT WAS REPORTED THE ADENOID BLADE BROKE. IT MAY HAVE BEEN DUE TO SCRUB TECH CLEANING. THERE WERE NO PATIENT CONSEQUENCES. SECOND OF 3 EVENTS; SEE 3007069406-2012-00430 AND 3007069406-2012-00432.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLASMABLADE TISSUE DISSECTION DEVICE GEI MEDTRONIC ADVANCED ENERGY LLC PLASMABLADE TNA 46645

Patients

Seq Age Sex Outcome Treatment
1 NI PULSAR GENERATOR