FDA Adverse Event Injury Summary report: N

OPENING JACK BACK ARM

MDR report key: 2822232 · Received May 23, 2012

Report

Report Number
1220246-2012-00093
Event Type
Injury
Date Received
May 23, 2012
Date of Event
April 27, 2012
Report Date
April 30, 2012
Manufacturer
ARTHREX, INC.
Product Code
LXH
PMA / PMN Number
EXEMPT
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PATIENT DEMOGRAPHICS (AGE AT TIME OF EVENT, DATE OF BIRTH, GENDER, WEIGHT) WERE REQUESTED BUT NOT PROVIDED. NO FURTHER PATIENT INFORMATION WAS PROVIDED AT THE TIME OF THIS REPORT OR MADE AVAILABLE IN RESPONSE TO FOLLOW-UP COMMUNICATION. NO ADDITIONAL ADVERSE CONSEQUENCES HAVE BEEN REPORTED FROM THIS EVENT. THIS DEVICE IS USED FOR TREATMENT. THE DEVICE WAS RECEIVED AND AN EVALUATION WAS CONDUCTED. DEVICE HISTORY RECORD REVEALED NOTHING RELEVANT TO THIS EVENT. COMPLAINT CONFIRMED. THE EVALUATION REVEALED THAT THE TWO RETURNED MATING PARTS, AN (B)(4), OPENING JACK TURN KEY, AND THE (B)(4), JACK BACK ARM, WERE LODGED TOGETHER AND COULD NOT BE TAKEN APART. THIS COMPLAINT IS STILL UNDER INVESTIGATION. AT THE CURRENT TIME THE CAUSE OF THE EVENT CANNOT BE DETERMINED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED FROM THE INVESTIGATION, A FOLLOW-UP REPORT WILL BE SUBMITTED. THE POTENTIAL CAUSE(S) OF THIS EVENT WILL BE COMMUNICATED TO THE EVENT REPORTER. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CASE WAS GOING SMOOTHLY UNTIL THE DOCTOR USED THE OSTEOTOMY JACK TO OPEN HIS CUT. THE JACK STOPPED OPENING OR CLOSING ONCE HE HAD THE CUT OPEN TO ABOUT 5MM. WITH NO OTHER WAY TO OPEN OSTEOTOMY, HE DECIDED HE COULD NOT USE THE IBALANCE IMPLANT. DOCTOR TRIED TO OPEN OSTEOTOMY WITH SMOOTH LAMINA SPREADERS, BUT THE TIBIA WAS TOO UNSTABLE AND WAS ROTATING. HE DECIDED TO PLATE WITH SYNTHES LOCKING PROXIMAL TIBIA PLATEAU AND COME BACK IN THREE MONTHS TO FINISH HTO WITH PUDDU/LOCKING PLATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPENING JACK BACK ARM ORTHOPEDIC MANUAL SURGICAL INSTRUMENT LXH ARTHREX, INC. 221126

Patients

Seq Age Sex Outcome Treatment
1 Other OPENING JACK TURN KEY: (B)(4)