FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2822151 · Received November 7, 2012

Report

Report Number
2531779-2012-13228
Event Type
Malfunction
Date Received
November 7, 2012
Report Date
September 24, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED (B)(4) 2012 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS INTACT AND ALL KEYS RESPONSIVE. REMOVED KEYPAD AND CONTAMINATION WAS NOTED UNDER "CONTRAST" AND "UP" KEY CONTACTS. THE BATTERY COMPARTMENT WAS CRACKED, BUT THE PUMP POWERED WITH THE RETURNED BATTERY CAP.

Description of Event or Problem · 1

THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED CONTAMINATION UNDER KEYPAD CONTACTS AND A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 13 YR