FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 2822026 · Received October 9, 2012

Report

Report Number
1824206-2012-06285
Event Type
Malfunction
Date Received
October 9, 2012
Date of Event
September 20, 2012
Report Date
September 20, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECH REPLACED THE SIDE RAIL END TUBES TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECH ALLEGED THAT THE SIDE RAIL WILL NOT LATCH. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1