FDA Adverse Event Malfunction Summary report: N

TOTALCARE BED

MDR report key: 2821993 · Received October 1, 2012

Report

Report Number
1824206-2012-06000
Event Type
Malfunction
Date Received
October 1, 2012
Date of Event
September 12, 2012
Report Date
September 12, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

HILL-ROM TECHNICIAN FOUND GROUND PRONG ON AC POWER CORD BROKEN INSIDE MOLDED END. HE REPLACED THE AC POWERCORD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

INFO RECEIVED INDICATED THE BED FAILED A GROUND TEST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TOTALCARE BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1900

Patients

Seq Age Sex Outcome Treatment
1