FDA Adverse Event Injury Summary report: N

GLOBAL ADVANTAGE CTA HD 44X18

MDR report key: 2821945 · Received November 7, 2012

Report

Report Number
1818910-2012-27417
Event Type
Injury
Date Received
November 7, 2012
Date of Event
October 24, 2012
Report Date
November 1, 2012
Manufacturer
DEPUY SUZHOU
Product Code
HSD
PMA / PMN Number
K000575
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

EXAMINATION OF THE RETURNED DEVICE SHOWS NOTHING OUTWARD TO SUGGEST PRODUCT PROBLEM. IT WAS INITIALLY REPORTED THE PATIENT FELL AFTER IMPLANTATION. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER SIMILAR REPORTS AGAINST THE PRODUCT/LOT CODE COMBINATIONS SINCE THEIR RELEASE TO DISTRIBUTION. REVIEW OF THE DEVICE HISTORY RECORDS FINDS NO RELATED MANUFACTURING DEVIATIONS OR ANOMALIES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN IDENTFIED.DEPUY CONSIDERS THIS INVESTIGATION CLOSED. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION THAT CHANGES THIS CONCLUSION BE RECEIVED, THE INVESTIGATION WILL BE REOPENED.

Description of Event or Problem · 1

THE PATIENT WAS REVISED DUE TO HEAD DISASSOCIATION ONLY THE HEAD WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GLOBAL ADVANTAGE CTA HD 44X18 DEPUY SHOULDER HEAD IMPLANT HSD DEPUY SUZHOU D11090267

Patients

Seq Age Sex Outcome Treatment
1 79 YR Required Intervention