- Report Number
- 1416980-2012-04238
- Event Type
- Malfunction
- Date Received
- November 7, 2012
- Date of Event
- September 18, 2012
- Report Date
- September 18, 2012
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915523
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). EVALUATION SUMMARY: THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS CONDITION IS AN ANCILLARY SERVICE EVENT OPENED DURING THE INVESTIGATION OF ANOTHER CONDITION. THE CAUSE WAS IDENTIFIED TO BE A MISSING P2 PUMP HEAD DOOR. THE DEVICE WILL BE RETURNED UNREPAIRED BECAUSE AN APPROVAL FOR THE REPAIR WAS NOT RECEIVED.
DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO BE MISSING A P2 PUMP HEAD DOOR. THIS WAS NOT REPORTED BY THE CUSTOMER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.