FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2821887 · Received November 7, 2012

Report

Report Number
1416980-2012-04238
Event Type
Malfunction
Date Received
November 7, 2012
Date of Event
September 18, 2012
Report Date
September 18, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS CONDITION IS AN ANCILLARY SERVICE EVENT OPENED DURING THE INVESTIGATION OF ANOTHER CONDITION. THE CAUSE WAS IDENTIFIED TO BE A MISSING P2 PUMP HEAD DOOR. THE DEVICE WILL BE RETURNED UNREPAIRED BECAUSE AN APPROVAL FOR THE REPAIR WAS NOT RECEIVED.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLOGARD INFUSION PUMP WAS FOUND TO BE MISSING A P2 PUMP HEAD DOOR. THIS WAS NOT REPORTED BY THE CUSTOMER. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1