NC TREK CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2012-06992
- Event Type
- Malfunction
- Date Received
- November 6, 2012
- Date of Event
- October 12, 2012
- Report Date
- October 12, 2012
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- K110134
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION AND THE REPORTED SEPARATION WAS CONFIRMED. BASED ON VISUAL ANALYSIS OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES. REVIEW OF THE COMPLAINT HISTORY OF THE REPORTED LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT DURING DEVICE PREPARATION, WHEN THE NC TREK DILATATION CATHETER WAS REMOVED FROM THE COILED HOOP, THE DISTAL BALLOON SEGMENT DETACHED FROM THE CATHETER. THE DEVICE WAS NOT USED AND THERE WAS NO PATIENT INVOLVEMENT. ANOTHER NC TREK WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER INCIDENT. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.