PERF CONS TIB BEARING 12MM S
Report
- Report Number
- 0009610576-2012-00024
- Event Type
- Injury
- Date Received
- November 6, 2012
- Date of Event
- October 1, 2012
- Report Date
- October 8, 2012
- Manufacturer
- BIOMET SPAIN, S.L.
- Product Code
- JWH
- PMA / PMN Number
- PK936274
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- PHYSICIAN
Narratives
THE REVIEW OF THE MANUFACTURING HISTORY RECORDS SHOWS THAT THE PRODUCT HAS BEEN MANUFACTURED ACCORDING TO BIOMET (B)(4)'S PRE-DEFINED SPECIFICATIONS. BIOMET (B)(4) IS THE SUB CONTRACTOR OF THIS PRODUCT. TEN PRODUCTS HAVE BEEN MANUFACTURED IN JUNE 2009 IN BIOMET (B)(4) FACILITY. THE ANALYSIS OF THE PRODUCT SHOWS THAT THE SCREW IS COMPLIANT WITH SPECIFICATION. THE INSERT IS DAMAGED SO IT WAS NOT POSSIBLE TO FULLY ANALYSE IT. THE SIZE OF THE IMPLANT IS COMPLIANT WITH THE REFERENCE AND DESIGNATION OF THE PRODUCT. THE MANUFACTURING HISTORY RECORD OF THE INLAY AND THE SCREW PACKAGED WITH THE INLAY SHOWS THAT THE PRODUCTS WERE MANUFACTURED ACCORDING TO BIOMET (B)(4) PREDEFINED SPECIFICATION. THE ANALYSIS OF THE PRODUCTS RETURNED (INLAY AND SCREW) DID NOT SHOW ANY ABNORMALITY. THE EXACT ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.
THE USER FACILITY IS OUTSIDE OF THE UNITED STATES. NO MEDWATCH REPORT WAS RECEIVED. ITEM WAS NOT RETURNED TO THE MANUFACTURER. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. NO FURTHER COMPLICATIONS HAVE BEEN REPORTED. (B)(4).
IT WAS REPORTED THAT PATIENT UNDERWENT KNEE PROCEDURE IN (B)(6) 2011. REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2012 DUE TO THE LOCKING SCREW COMING LOOSE. SCREW WAS REMOVED AT TIME OF REVISION. NO FURTHER INFORMATION HAS BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERF CONS TIB BEARING 12MM S | PROSTHESIS, KNEE | JWH | BIOMET SPAIN, S.L. | N/A | 0000446888 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |