PULSAR GENERATOR
Report
- Report Number
- 3007069406-2012-00356
- Event Type
- Malfunction
- Date Received
- October 25, 2012
- Report Date
- January 3, 2012
- Manufacturer
- MEDTRONIC ADVANCED ENERGY LLC
- Product Code
- GEI
- PMA / PMN Number
- K082786
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS MDR IS BEING FIELD BASED ON A RETROSPECTIVE REVIEW OF COMPLAINTS RECEIVED BY THE MANUFACTURER FOR THE TIME PERIOD (B)(4) 2010 THROUGH (B)(4) 2012. THE PULSAR GENERATOR WAS RECEIVED IN POOR CONDITION, WITH STAINS, WORN RUBBER FEET AND OTHER IMPERFECTIONS. THE UNIT DELIVERED RF ENERGY CORRECTLY INTO THE FIXED RESISTORS. THE UNIT GAVE E3 ERRORS DURING TP-005 TESTING. THE C15 WAS REPLACED ON THE RF CONTROLLER PCBA AND THE UNIT NO LONGER DISPLAYED E3 DURING THE IMBALANCE TESTING. THE UNIT WAS REPAIRED. APPLICABLE SOFTWARE AND HARDWARE UPGRADES WERE PERFORMED. THE UNIT PASSED FINAL TESTING AND WAS RECERTIFIED FOR USE.
IT WAS REPORTED THE GENERATOR WAS NOT WORKING, INTERMITTENT DURING A CASE. THEY SWITCHED TO A FINGER CONTROL AND WERE ABLE TO CONTINUE WITH THE CASE. THERE WERE NO PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PULSAR GENERATOR | GEI | MEDTRONIC ADVANCED ENERGY LLC | PULSAR | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI | PLASMABLADE |