FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2821161 · Received October 25, 2012

Report

Report Number
1627487-2012-02580
Event Type
Injury
Date Received
October 25, 2012
Date of Event
October 3, 2012
Report Date
October 3, 2012
Manufacturer
ST. JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REFERENCE MFR REPORT: 1627487-2012-02579. IT WAS REPORTED THE PT FELT A BURNING SENSATION AT THE IPG SITE WHILE CHARGING. SHE REPORTED SHE ONLY NEEDED TO CHARGE ABOUT 15 MINS AND THE WARMTH WAS TOLERABLE. IT WAS REPORTED THE IPG SITE DID NOT APPEAR RED OR IRRITATED. THE SJM REP ADVISED THE PT OF CHARGING PRECAUTIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST. JUDE MEDICAL - NEUROMODULATION 3721 3772654

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention IMPLANT DATE:| SCS LEAD: MODEL 3186 (2)| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2)