FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2820883
·
Received November 6, 2012
Report
- Report Number
- 2531779-2012-13216
- Event Type
- Malfunction
- Date Received
- November 6, 2012
- Report Date
- October 10, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD RUBBER WAS INTACT. EVALUATION REVEALED THAT THE CONTRAST, DOWN AND UP BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND THE OK BUTTON RESPONDED APPROPRIATELY. THERE WAS CONTAMINATION PRESENT UNDER ALL THE KEY CONTACTS.
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE CONTRAST, UP AND DOWN BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER ALL OF THE KEYS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2012.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR |