FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2820883 · Received November 6, 2012

Report

Report Number
2531779-2012-13216
Event Type
Malfunction
Date Received
November 6, 2012
Report Date
October 10, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD RUBBER WAS INTACT. EVALUATION REVEALED THAT THE CONTRAST, DOWN AND UP BUTTONS WERE INTERMITTENTLY UNRESPONSIVE AND THE OK BUTTON RESPONDED APPROPRIATELY. THERE WAS CONTAMINATION PRESENT UNDER ALL THE KEY CONTACTS.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE CONTRAST, UP AND DOWN BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER ALL OF THE KEYS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 52 YR