ROD 06 SOFT L100 TI
Report
- Report Number
- 8030965-2012-01167
- Event Type
- Injury
- Date Received
- October 31, 2012
- Date of Event
- September 24, 2012
- Report Date
- October 24, 2012
- Manufacturer
- SYNTHES GMBH
- Product Code
- HSB
- PMA / PMN Number
- K992739
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
(B)(4): INVESTIGATION COORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES AT S1 BOTH RODS SLIPPED AND WERE HITTING AGAINST THE LOCKING CAP. IT IS VISIBLE THAT THE RODS HAD TO BE HEAVILY BENT IN THE SAGITTAL PLANE TO FIT THE IMPLANTS (TRACES ON ROD INDICATE BEND). ACCORDING TO THE TRACES ON THE RODS AND IN THE SADDLE, AT LEAST THE IMPLANT WITH THE LOCKING CAP 2 MIGHT HAVE NOT BEEN WELL ALIGNED WITH THE ROD, WHICH MIGHT EXPLAIN THE SLIPPAGE OF THE CORRESPONDING ROD. WHEN ONE ROD SLIPS, THERE IS A RISK FOR THE SECOND ROD TO SLIP FROM OVERLOAD. THE RETURNED DEVICE WAS ANALYZED FOR CONFORMANCE TO SPECIFICATIONS AND A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED. THERE WERE NO ABNORMAL FINDINGS. NO PRODUCT FAULT COULD BE DETECTED. (B)(4): PLACEHOLDER.
SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. THE MFG DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.
A DEVICE REPORT FROM SYNTHES (B)(6) PROVIDES INFO FROM A FACILITY IN (B)(4) REGARDING A REVISION PROCEDURE (UNK DATE); ARTHRODESIS, REQUIRED DUE TO LOOSENING OF ROD SCREWS, BILATERALLY AT S1. THE RODS LEFT AND RIGHT DROPPED/SLID OUT OF THE HEAD (LOCKING CAPS WERE STILL ATTACHED). THERE WAS NO BREAKAGE OF THE IMPLANT. METALLOSIS PRESENCE OF BILATERAL REGION AND PSEUDOARTHROSIS WERE NOTED. THIS IS REPORT #3 OF 6 FOR THE SAME EVENT.
THE PATIENT UNDERWENT A REVISION ARTHRODESIS, AS THE LOCKING CAP ON SCREWS S1 (BILATERAL) CAME LOOSE. THE RODS LEFT AND RIGHT SLID OUT OF THE HEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ROD 06 SOFT L100 TI | ROD | HSB | SYNTHES GMBH | 3789701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Required Intervention |