FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2820626 · Received November 2, 2012

Report

Report Number
1720753-2012-09029
Event Type
Malfunction
Date Received
November 2, 2012
Date of Event
September 7, 2012
Report Date
November 2, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INTERCONNECT CABLE WAS NOT WORKING AND IT PREVENTED THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF PATIENT INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1