FDA Adverse Event Injury Summary report: N

DRAEGER ISOLETTE INFNT INCUBATOR

MDR report key: 2820609 · Received October 31, 2012

Report

Report Number
2510954-2012-00004
Event Type
Injury
Date Received
October 31, 2012
Date of Event
September 17, 2012
Report Date
September 27, 2012
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC.
Product Code
FMZ
PMA / PMN Number
K960980
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A DRAEGER SVC REP WAS REQUESTED BY THE FACILITY TO CHECK THE HUMIDITY MODULE FOR PROPER OPERATION AND THE SERVICE REPORT INDICATED THE MODULE MET MANUFACTURE SPECIFICATIONS AND FUNCTIONED AS INTENDED. IN THIS CASE THE USER HAND WAS BURNED FROM STEAM VAPORS WHEN ATTEMPTING TO REFILL THE RESERVOIR. THE INSTRUCTIONS FOR USE (IFU) SECTION "FILLING THE HUMIDITY RESERVOIR" PROVIDES THE USER WITH INSTRUCTIONS AND ILLUSTRATIONS REGARDING REFILLING THE RESERVOIR DURING USE. THE EVAPORATOR AND VAPOR PORT ARE LOCATED AT THE REAR OF THE RESERVOIR ASSEMBLY. REFILLING THE RESERVOIR DOES NOT REQUIRE COMPLETE REMOVAL FROM THE UNIT. BASED ON THIS, IT IS NOT KNOWN HOW THE USER WAS EXPOSED TO THE STEAM VAPORS WITHOUT COMPLETELY REMOVING THE RESERVOIR. A WARMING IN THE IFU STATES "THE EVAPORATOR CAN BE SUFFICIENTLY HOT TO CAUSE BURNS. AVOID REMOVING OR TOUCHING THE EVAPORATOR UNTIL THE HUMIDITY RESERVOIR HAS BEEN DISCONNECTED FROM THE INCUBATOR FOR AT LEAST 45 MINS. NONCOMPLIANCE MAY RESULT IN DEATH OR SERIOUS INJURY." THERE ARE ALSO THE INTERNATIONAL SYMBOLS FOR "CAUTION" AND "CAUTION: HOT SURFACE" EMBOSSED ON THE TOP SURFACE OF THE RESERVOIR ASSEMBLY AS A VISUAL WARNING TO THE USER. DRAEGER HAS MADE REPEATED ATTEMPTS TO OBTAIN ADDITIONAL INFO FROM THE FACILITY IN REGARD TO THE EXTENT OF INJURY TO THE USER AND MORE DETAIL AS TO WHAT TRANSPIRED TO CAUSE THE INJURY TO HAVE OCCURRED. DRAEGER HAS NOT RECEIVED A RESPONSE TO THESE REQUESTS AFTER MAKING MULTIPLE ATTEMPTS. WITH THE LIMITED INFO PROVIDED NO CONCLUSIONS CAN BE DRAWN.

Description of Event or Problem · 1

A NURSE WAS BURNED/SCALDED WHEN PULLING OUT THE RESERVOIR DRAWER TO FILL THE WATER RESERVOIR. IT WAS ALSO REPORTED THAT THERE WAS A RELEASE OF STEAM WHEN THE DRAWER WAS PULLING OUT. AS OF (B)(6) 2012, THE NURSE WAS STILL OUT OF WORK DUE TO THE INJURY. THE CUSTOMER REQUESTED TO HAVE A DRAEGER SERVICE ENGINEER COME TO THE HOSPITAL TO CHECK OUT THE ISOLETTE 8000 FOR PROPER FUNCTION. AFTER MULTIPLE ATTEMPTS, NO ADDITIONAL DETAILS COULD BE OBTAINED FROM THE HOSPITAL. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DRAEGER ISOLETTE INFNT INCUBATOR INFANT INCUBATOR FMZ DRAEGER MEDICAL SYSTEMS, INC. ISOLETTE 8000 NA

Patients

Seq Age Sex Outcome Treatment
1 Other