FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 2820464 · Received October 31, 2012

Report

Report Number
1720753-2012-08958
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 18, 2012
Report Date
October 31, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE KVP CONTROLLER BOARD AND THE INTERFACE BOARD. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN SITUATION DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURRED. THIS COULD RESULT IN ADDITIONAL UNNECESSARY DELAY AND/OR ANESTHESIA. THERE IS NO REPORT OF PATIENT INJURY.