FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2820439 · Received October 31, 2012

Report

Report Number
1720753-2012-08944
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 15, 2012
Report Date
October 31, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND COULD NOT REPRODUCE THE REPORTED PROBLEM. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WOULD NOT SAVE IMAGES. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1