ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2820419
·
Received November 6, 2012
Report
- Report Number
- 2531779-2012-13207
- Event Type
- Malfunction
- Date Received
- November 6, 2012
- Report Date
- October 10, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD SYMBOLS WERE WORN BUT THE KEYPAD RUBBER WAS INTACT. EVALUATION REVEALED THAT THE CONTRAST BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE PRESSES TO ELICIT A RESPONSE AND THE OTHER BUTTONS RESPONDED APPROPRIATELY. THERE WAS CONTAMINATION PRESENT UNDER THE CONTACTS OF ALL BUTTONS.
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE ALL OF THE KEY CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.