FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2820399 · Received October 31, 2012

Report

Report Number
1720753-2012-08921
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 18, 2012
Report Date
October 31, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE PS1 AND PS2 POWER SUPPLY VOLTAGES WERE ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM KEPT SHUTTING DOWN ON IT'S OWN. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1