FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2820311 · Received October 31, 2012

Report

Report Number
8020893-2012-01102
Event Type
Malfunction
Date Received
October 31, 2012
Date of Event
October 5, 2012
Report Date
October 9, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
RI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR HAD A BLANK SCREEN. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND REPLACED THE POWER SUPPLY ASSEMBLY. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1