FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 2820226
·
Received October 31, 2012
Report
- Report Number
- 2518422-2012-02171
- Event Type
- Malfunction
- Date Received
- October 31, 2012
- Date of Event
- October 2, 2012
- Report Date
- October 2, 2012
- Manufacturer
- RESPIRONICS INC.
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AU
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, ISSUES RELATED TO THE SENSOR BOARD AND INVERTER BOARD WERE OBSERVED. THE DEVICE'S INVERTER BOARD WERE REPLACED TO ADDRESS THE ISSUE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR'S DISPLAY WAS DEFECTIVE. NO HARM OR INJURY WAS REPORTED.