FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2820079 · Received November 6, 2012

Report

Report Number
2531779-2012-13156
Event Type
Malfunction
Date Received
November 6, 2012
Date of Event
October 8, 2012
Report Date
October 8, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NM, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH COULD NOT BE DUPLICATED DURING TESTING. THE FORCE SENSOR WAS FOUND TO BE OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT WAS OUT OF SPECIFICATION. THERE WAS EVIDENCE OF CONTAMINATION OBSERVED ON THE FORCE SENSOR PLATE, AND THERE WAS MOISTURE CORROSION DAMAGE OBSERVED INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6). (B)(4).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR RECURRING LOSS OF PRIMES. INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT WAS OUT OF SPECIFICATION. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(6) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 26 YR