FDA Adverse Event Summary report: N

DIRECT SHIELDED DOOR

MDR report key: 28198 · Received October 11, 1995

Report

Report Number
MW4000920
Date Received
October 11, 1995
Date of Event
May 2, 1995
Report Date
May 17, 1995
Manufacturer
PPB, INC.
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
AZ, US
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

DOOR JAMMED IN PARTIALLY-OPEN POSITION. PT IN ROOM AND RPTR WAS NOT ABLE TO FREE PT UNTIL JACK OBTAINED TO PRY DOOR OPEN. LATER IT WAS DETERMINED THAT WHEEL HAD BROKEN AND JAMMED ON DOOR PLATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIRECT SHIELDED DOOR DIRECT SHIELDED DOOR PPB, INC. CC21

Patients

Seq Age Sex Outcome Treatment
1 *