ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2012-13126
- Event Type
- Malfunction
- Date Received
- November 4, 2012
- Report Date
- October 9, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW UP #1 SUBMITTED: (B)(4) 2012 - DEVICE EVALUATION: A RETAINED CARTRIDGE SAMPLE FROM LOT # B201876 WAS EVALUATED. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST, FILL TEST, AND FORCE TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE (OCCLUSION AND LOSS OF PRIME), CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.
THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
THE PATIENT CLAIMED THERE ARE AIR BUBBLES IN THE CARTRIDGE POSSIBLY DUE TO A LOOSE LUER LOCK. THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE AIR BUBBLES IN THE CARTRIDGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM IR1200/1250/2020/OTP CART | B201876 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |